Which of the following compounds is used as a smoke source for the smoke pattern test of BSCs
- Tetrachloromethane
- Fluorine tetrachloride
- Titanium tetrachloride
- Tetrachlorethylene
To filter a minimum of 0.3μm particles, how efficient should a HEPA filter be?
- 99.13
- 99.39
- 99.69
- 99.99
Which of the following actions when operating BSCs is wrong?
- After use, empty the cabinet.
- Before use, turn on the BSC for a while.
- Turn off the BSC immediately after use and take out the items.
- Wipe workbench with alcohol.
Among the following National Sanitation Foundation (NSF) standards, which one is the standard for BSCs?
- NSF29
- NSF39
- NSF49
- NSF59
When testing inflow velocity, a minimum of how many readings shall be measured at the sash?
- 5
Among the following Class II BSCs, which type has a face velocity of 75 ft/m?
- A1
- A2
- B1
- B2
When testing downflow velocity, a minimum of how many readings across the workbench shall be measured in a horizontal plane 10 cm above the bottom edge of the sash?
- 2X8
- 3X7
- 4X6
- 5X7
Among the following Class II BSCs, which one does not circulate airflow in the work area?
- A1
- A2
- B1
- B2
Which of the following chemicals can be used for the HEPA filter leak test?
- PAO
- SiO2
- H2O2
- MgO
Which of the following information is not included in a regular BSC test report?
- HEPA filter leak test
- Cabinet integrity test
- Inflow velocity test